A Monte Carlo Method to Estimate Radiation Dose from Cone Beam Computed Tomography Public Deposited

Downloadable Content

Download PDF
Last Modified
  • March 19, 2019
Creator
  • Dunn, Jacob S.
    • Affiliation: School of Dentistry, Oral Pathology Section, Oral and Maxillofacial Radiology Graduate Program
Abstract
  • Objectives: This study compares effective dose determination of four large field of view CBCT units (NewTom-3G, Galileos Comfort Plus, CS-9300, iCat-FLX,) using a Monte Carlo software analysis method (PCXMC) and dosimetry using anthropomorphic phantoms. Methods: Previous research provided phantom effective dose comparisons. Field-of-view and phantom positioning were duplicated in the software. Software and phantom dosimetry values were compared. Descriptive statistics, chi-square, and logistic regression, were used to analyze the data. The null-hypothesis that there is no statistically significant difference between the dosimetry values of the anthropomorphic phantom and the calculated values of the PCXMC software was tested. Results: PCXMC simulated scans produced dose values within 20% of the phantom dosimetry 48-58% of the time. Conclusions: While the software calculations are simpler to perform than phantom dosimetry, imprecise calculated results make this program less effective for CBCT dosimetry in dentistry.
Date of publication
Keyword
Subject
DOI
Identifier
Resource type
Rights statement
  • In Copyright
Advisor
  • Ludlow, John B.
  • Mol, André
  • Ivanovic, Marija
Degree
  • Master of Science
Degree granting institution
  • University of North Carolina at Chapel Hill Graduate School
Graduation year
  • 2015
Language
Publisher
Place of publication
  • Chapel Hill, NC
Access
  • There are no restrictions to this item.
Parents:

This work has no parents.

Items