A Design and Performance Analysis of NIEHS Laboratory Hoods Public Deposited

Downloadable Content

Download PDF
Last Modified
  • February 28, 2019
Creator
  • Smith, Thomas C.
    • Affiliation: Gillings School of Global Public Health, Department of Environmental Sciences and Engineering
Abstract
  • The laboratory hoods found at the National Institute of Environmental Health Sciences were quantitatively tested and evaluated in terms of containment performance. The hoods were challenged with a tracer gas. Sulfur Hexafluoride, and peak concentrations leaking from the hood were measured in the breathing zone of a mannequin. The tests proved that laboratory hood performance is subject to baffle position, face velocity, tracer gas challenge height and bottom slot obstructions. These factors had statistically significant effects, P <0.0001, on overall containment efficiency. The results helped to identify the parameters which could be manipulated by the employee, the Health and Safety Branch and facility engineers to provide optimum performance and reduction of hood leakage to below 0.1 ppm.
Date of publication
Resource type
Rights statement
  • In Copyright
Advisor
  • Reist, Parker C.
  • Dement, John M.
  • Flynn, Michael
Degree
  • Master of Science in Environmental Engineering
Academic concentration
  • Industrial Hygiene
Degree granting institution
  • University of North Carolina at Chapel Hill
Graduation year
  • 1990
Language
Deposit record
  • ad7e9272-7c4f-43e0-afeb-7a5f2526ad76
Parents:

This work has no parents.

Items